Validation of a discovery MI 4-ring model according to the NEMA NU 2-2018 standards: from Monte Carlo simulations to clinical-like reconstructions.

Fiche publication


Date publication

janvier 2024

Journal

EJNMMI physics

Auteurs

Membres identifiés du Cancéropôle Est :
Pr COCHET Alexandre


Tous les auteurs :
Merlet A, Presles B, Su KH, Salvadori J, Sayah F, Jozi H, Cochet A, Vrigneaud JM

Résumé

We propose a comprehensive evaluation of a Discovery MI 4-ring (DMI) model, using a Monte Carlo simulator (GATE) and a clinical reconstruction software package (PET toolbox). The following performance characteristics were compared with actual measurements according to NEMA NU 2-2018 guidelines: system sensitivity, count losses and scatter fraction (SF), coincidence time resolution (CTR), spatial resolution (SR), and image quality (IQ). For SR and IQ tests, reconstruction of time-of-flight (TOF) simulated data was performed using the manufacturer's reconstruction software.

Mots clés

GATE, Monte Carlo simulation, NEMA, Nuclear medicine, PET

Référence

EJNMMI Phys. 2024 01 31;11(1):13