Thyroid-stimulating hormone and free thyroxine on the ADVIA Centaur immunoassay system: a multicenter assessment of analytical performance.

Fiche publication


Date publication

septembre 2013

Auteurs

Membres identifiés du Cancéropôle Est :
Dr AGIN Arnaud, Dr HEURTAULT Béatrice


Tous les auteurs :
Reix N, Massart C, d'Herbomez M, Gasser F, Heurtault B, Agin A

Résumé

OBJECTIVES: We assessed the analytical performance of the TSH and FT4 assays on ADVIA Centaur in a multicenter national evaluation. DESIGN AND METHODS: A precision study and a method comparison were performed. Reference values stated by the manufacturer were checked from 379 normal subjects. RESULTS: For TSH and FT4, the intra-assay CVs were below 2.3 and 5.2%, respectively, and the inter-assay CVs below 4.4% and 7.2%, respectively. Therefore, the precision and reproducibility were acceptable. Bland-Altman bias plots revealed good correlation and agreement with Cobas assays. TSH and FT4 data yielded reference ranges of 0.64-3.24 mIU/L and 10.5-18.9 pmol/L, respectively. CONCLUSION: These assays demonstrate reliable characteristics. The reference ranges obtained can be used for interpretation of thyroid function.

Référence

Clin Biochem. 2013 Sep;46(13-14):1305-8