Adaptive step size LMS improves ECG detection during MRI at 1.5 and 3 T.

Fiche publication


Date publication

décembre 2017

Journal

Magma (New York, N.Y.)

Auteurs

Membres identifiés du Cancéropôle Est :
Pr FELBLINGER Jacques


Tous les auteurs :
Guillou A, Sellal JM, Ménétré S, Petitmangin G, Felblinger J, Bonnemains L

Résumé

We describe a new real-time filter to reduce artefacts on electrocardiogram (ECG) due to magnetic field gradients during MRI. The proposed filter is a least mean square (LMS) filter able to continuously adapt its step size according to the gradient signal of the ongoing MRI acquisition.

Mots clés

Electric artefact, Magnetic field gradient, Noise reduction

Référence

MAGMA. 2017 Dec;30(6):567-577